Doble Lemke LDP-5 Hand-held PD Detector
 |
Applications:
- Detection of dangerous partial discharges (PD)
- Mode for differential measurement with respect to PD fault location
- Diagnostics on HV apparatus in field (for instance cable sealing ends and joints, surge arrestors, power transformers, voltage transformers and electrical machines)
- Evaluation of the PD behaviour of low voltage components (for instance in the case of opto-couplers, capacitors, transformers and motors)
|
User Benefits
- Battery powered portable device for both, on-site and laboratory PD diagnosis
- Low Cost PD detector for research and development
Specifications:
- Detection of the electromagnetic transients of PD events
- Measuring the PD quantity "apparent charge"
- Exchangeable capacitive and inductive sensors
- Potentialfree signal transmission via fibre optic link
- Noise reduction mode
- Measuring sensitivity if charges pulses are directly injected : 0,05 pC
- Gain steps : 1, 2, 5, 10
- Vernier range : 1 : 5
- Output (CHARGE) at full load
- Pulse shape : triangular, positive
- Pulse magnitude : 4 V approx.
- Pulse rise time : < 2 µs
- Pulse duration : 50 µs approx.
- Noise level : < 0,2 V
- Pulse resolution performance for superposition error < 30%
- Pulse repetition rate : < 100 kHz
- Double pulse distance : > 2 µs
- Power supply : 8 Mignon batteries (LR6)
- Dimensions : 220 × 60 × 100
- Weight : 1 kg approx.
|
|
|
|
|
|
|
|
|
|
|
Brochures:
LDP-5
Accessories:
LDC-5 - Calibrator for external calibration of the PD measuring circuit in pC of apparent charge up to 500 pC
LDA-5/U4 - UHF-Converter
LDO-5 - Optical receiver for potential-free signal transmission
LDA-5/S4 - Ultra Sonic Adapter
LDA-5/K - Acoustic Converter
Carry case "G" which contains the basic equipment
Carry case "E" which contains the basic equipment and main accessories
Carry case "B" which contains extended accessories
|
|